FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3940404 · Received April 17, 2014

Report

Report Number
3008642652-2014-01158
Event Type
Malfunction
Date Received
April 17, 2014
Date of Event
March 17, 2014
Report Date
April 11, 2014
Manufacturer
ZOLL LIFECOR CORP
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BED SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED CABLE) HAS BEEN CONFIRMED. UPON INVESTIGATION THE CABLE CONNECTING THE DISTRIBUTION NODE TO THE REAR THERAPY ELECTRODES WAS PULLED FROM THE STRAIN RELIEF, DAMAGING WIRES. THE ROOT CAUSE FOR THE STRAINED CABLE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE STRAINED CABLE. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A ZOLL PT SERVICE REPRESENTATIVE (PSR) CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT A (B)(6) MALE PT HAD A DAMAGED ELECTRODE BELT CABLE. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235612 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORP WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 69 YR