FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3940404
·
Received April 17, 2014
Report
- Report Number
- 3008642652-2014-01158
- Event Type
- Malfunction
- Date Received
- April 17, 2014
- Date of Event
- March 17, 2014
- Report Date
- April 11, 2014
- Manufacturer
- ZOLL LIFECOR CORP
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BED SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED CABLE) HAS BEEN CONFIRMED. UPON INVESTIGATION THE CABLE CONNECTING THE DISTRIBUTION NODE TO THE REAR THERAPY ELECTRODES WAS PULLED FROM THE STRAIN RELIEF, DAMAGING WIRES. THE ROOT CAUSE FOR THE STRAINED CABLE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE STRAINED CABLE. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A ZOLL PT SERVICE REPRESENTATIVE (PSR) CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT A (B)(6) MALE PT HAD A DAMAGED ELECTRODE BELT CABLE. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 235612 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORP | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR |