LIEFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2014-01130
- Event Type
- Malfunction
- Date Received
- April 17, 2014
- Date of Event
- February 13, 2014
- Report Date
- April 9, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CODE 102) HAS BEEN CONFIRMED. AS RECEIVED, THE MONITOR FAILED INCOMING TESTING AND DISPLAYED SVC CODE 102. THE CAUSE FOR THE CODE 102 AND INCOMING TEST FAILURE IS AN OPEN RESISTOR R45 ON THE COMPUTER/ANALOG (CA) BOARD. THE CAUSE FOR THE OPEN RESISTOR IS EXCESSIVE CURRENT. THE CAUSE FOR THE EXCESSIVE IS BROKEN LEADS ON HIGH-VOLTAGE CAPACITORS C21 AND C22 ON THE DEFIBRILLATOR BOARD. THE ROOT CAUSE FOR THE BROKEN LEADS ON C21 AND C22 COULD NOT BE POSITIVELY IDENTIFIED, BUT THE DAMAGE LIKELY RESULTED FROM THE MONITOR IMPACTING A HARD SURFACE. ROOT CAUSE ANALYSIS OF SIMILAR EVENTS FOUND THAT EXCESSIVE STRAIN ON THE PRINTED CIRCUIT ASSEMBLY (PCA) FROM SEVERE IMPACT CAN CAUSE FRACTURED SOLDER CONNECTIONS. A MECHANICAL DESIGN CHANGE (SEE PMA SUPPLEMENT S039) TO REDUCE THE PCA STRAIN WAS APPROVED BY FDA ON 12/20/2012. IMPLEMENTATION BEGAN ON 01/21/2013. RESULTS WILL BE MONITORED. NO ADVERSE EVENT RESULTED FROM THE OPEN RESISTOR. THE PATIENT RECEIVED A REPLACEMENT MONITOR.
ZOLL CUSTOMER SUPPORT CONTACTED A (B)(6) MALE PT TO EXCHANGE HIS MONITOR. THE PT'S DOWNLOADED DATA REVEALED MULTIPLE OCCURRENCES OF SVC CODE 102. THE PT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 235610 | LIEFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |