FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3940379
·
Received April 17, 2014
Report
- Report Number
- 3008642652-2014-01101
- Event Type
- Malfunction
- Date Received
- April 17, 2014
- Date of Event
- December 16, 2013
- Report Date
- April 7, 2014
- Manufacturer
- ZOLL LIFECORCORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WON'T POWER UP) HAS BEEN CONFIRMED. UPON INVESTIGATION, THE BATTERY CHARGER/MODEM WAS UNABLE TO POWER UP. THE CAUSE FOR THE FAILURE WAS ISOLATED TO A DEFECTIVE POWER SUPPLY BRICK. THE ROOT CAUSE FOR THE DEFECTIVE POWER SUPPLY COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM. THE PATIENT RECEIVED A REPLACEMENT BATTERY CHARGER/MODEM.
Description of Event or Problem · 1
A (B)(4) DISTRIBUTOR CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT A (B)(6) PT'S BATTERY CHARGER/MODEM WASN'T POWERING UP. THE PT WAS ISSUED A REPLACEMENT BATTERY CHARGER/MODEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 235609 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECORCORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |