FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 3000 SYSTEM

MDR report key: 3940376 · Received April 17, 2014

Report

Report Number
3008642652-2014-01148
Event Type
Malfunction
Date Received
April 17, 2014
Date of Event
March 27, 2014
Report Date
April 10, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. AS RECEIVED, THE TRUNK CABLE CONNECTOR HOUSING WAS BROKEN. THE CONNECTOR LOCKING NUT WAS DAMAGED WHICH WOULD PREVENT THE ELECTRODE BELT FROM PROPERLY SECURING INTO THE MONITOR. THE ROOT CAUSE FOR THE DAMAGED LOCKING NUT CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY PHYSICAL ABUSE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CONNECTOR. PATIENT WAS RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

DURING INVESTIGATION INTO A (B)(6) FEMALE PT'S ELECTRODE BELT, WHICH WAS RETURNED FOR UNRELATED ISSUE, A REPORTABLE PROBLEM WAS DISCOVERED. THE ELECTRODE BELT CONNECTOR LOCKING NUT WAS CRACKED. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235604 LIFEVEST WCD 3000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 3000 NA

Patients

Seq Age Sex Outcome Treatment
1 67 YR