LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2014-01165
- Event Type
- Malfunction
- Date Received
- April 17, 2014
- Date of Event
- March 27, 2014
- Report Date
- April 14, 2014
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVAL SUMMARY: DEVICE EVAL OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CODE 107) HAS BEEN CONFIRMED. AS RECEIVED, THE MONITOR WAS ABLE TO PROPERLY DETECT AND TREAT. HOWEVER, UPON EVAL, THE MONITOR WAS EXPERIENCING ABNORMAL SHUTDOWNS WAS LIKELY AN INTERMITTENT CONNECTION AT BGA COMPONENT U1003 (PLD) ON CA BOARD SN (B)(4). THE SUSPECT INTERMITTENT CONNECTION IS LIKELY DUE TO A FRACTURED BGA SOLDER CONNECTION INDUCED BY MECHANICAL STRESS. THE EXACT SOURCE OF THE MECHANICAL STRESS HAS NOT BEEN POSITIVELY IDENTIFIED BUT THE FRACTURED CONNECTION MAY HAVE BEEN ACCELERATED THROUGH ROUGH HANDLING. A ROOT CAUSE INVESTIGATION IS CURRENTLY UNDERWAY. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE MONITOR. THE PT RECEIVED A REPLACEMENT MONITOR.
A (B)(6) FEMALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT A SERVICE CODE 107. THE PT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 235937 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR |