ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-20383
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Report Date
- July 8, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP # 1 DATE OF SUBMISSION 09/10/2014 ¿ DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP¿S BLACK BOX HISTORY SHOWED EVIDENCE OF POWER REBOOTS OCCURRING. VISUAL INSPECTION REVEALED THAT THERE WERE BATTERY COMPARTMENT CRACKS IN THE THREAD AREA AND BELOW THE GRIP PAD. DURING TESTING, THE PUMP POWERED ON APPROPRIATELY WITH THE RETURNED BATTERY CAP. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO POWER REBOOTS, POWER LOSSES, OR CALL SERVICE ALARMS OCCURRING. THE PUMP CASE WAS REMOVED AND NO EVIDENCE OF MOISTURE CONTAMINATION WAS FOUND INSIDE THE PUMP. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED THAT THE DISPLAY WAS DIM AND DISCOLORED. THE COMPLAINT THAT THE PUMP HAD NO POWER WAS OBSERVED IN THE PUMP HISTORY BUT WAS NOT ABLE TO BE DUPLICATED DURING INVESTIGATION.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (NO POWER) ISSUE. IT WAS REPORTED THAT THE PUMP HAD NO POWER. IT WAS NOTED THAT THE POWER ISSUE OCCURRED DURING OR IMMEDIATELY AFTER A BATTERY CHANGE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 418137 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR |