SUNQUEST COPATHPLUS
Report
- Report Number
- 2029302-2014-00001
- Event Type
- Malfunction
- Date Received
- March 6, 2014
- Date of Event
- February 10, 2014
- Report Date
- March 5, 2014
- Manufacturer
- SUNQUEST INFORMATION SYSTEM
- Product Code
- NVV
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
FOURTEEN SITES HAVE BEEN NOTIFIED VIA PRODUCT SAFETY NOTICE (PSN) AND A CORRECTION WILL BE AVAILABLE IN COPATHPLUS VERSION 6.1.1 AND ABOVE. THERE ARE 0 SITES THAT HAVE ALREADY BEEN CORRECTED. STEPS TO REPRODUCE: LOGON AS SUNQUEST USER AND ACCESSION A SURGICAL SPECIMEN. NAVIGATE TO THE ELECTRONIC SIGNOUT FOR FINAL REPORT WINDOW FOR THE SPECIMEN. ENTER A LARGE AMOUNT OF TEXT IN THE FINAL DIAGNOSIS FIELD AND THEN CLICK ON SAVE/QUICK SIGN/OUT TO OPEN THE QUICK SIGN OUT REPORT. ADD SOME TEXT TO THE FIRST PARAGRAPH OF THE FINAL DIAGNOSIS TEXTFIELD AND CLICK CTRL-Z TO UNDO THESE ADDITIONS. CLICK ON EDIT SPECIMEN AND THEN ON OK TO SAVE CHANGES. ON RETURNING TO THE ELECTRONIC SIGNOUT FOR FINAL REPORT NOTICE THAT THE FIRST PARAGRAPH OF THE DIAGNOSIS IS MISSING. CLICK ON SAVE/QUICK SIGN/OUT TO OPEN THE QUICK SIGN OUT REPORT. CLICK CTR-O AND A SYSTEM ERROR WINDOW IS DISPLAYED. IF THE USER CLICK OK THEN THE COPATH APPLICATION CLOSES DOWN.
WHILE USING THE COPATHPLUS QUICK SIGN-OUT REPORTS FEATURE, IF A USER PRESSES THE KEYBOARD SHORTCUT "CRL-Z" WHILE EDITING TEXT IN A DIAGNOSTIC TEXT FIELD, PORTIONS OF PREVIOUSLY SAVED TEXT MAY DISPPEAR UPON SAVING THE REPORT. WHILE USING THE COPATHPLUS QUICK SIGN-OUT REPORTS FEATURE, IF A USER PRESSES THE KEYPAD SHORTCUT "CTRL O" THE APPLICATION MAY CLOSE UNEXPECTEDLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 134202 | SUNQUEST COPATHPLUS | COPATHPLUS | NVV | SUNQUEST INFORMATION SYSTEM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |