FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3936936 · Received July 15, 2014

Report

Report Number
2531779-2014-20207
Event Type
Malfunction
Date Received
July 15, 2014
Report Date
July 7, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

PRODUCT ANALYSIS: THE PUMP WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/19/2014 WITH THE FOLLOWING FINDINGS: MOISTURE WAS OBSERVED WITHIN THE BATTERY COMPARTMENT. NO DAMAGE WAS OBSERVED TO THE PUMP¿S BATTERY COMPARTMENT OR BATTERY CAP. THE BATTERY CAP WAS ABLE TO PROPERLY SECURE TO THE PUMP. THE PUMP PASSED LEAK TESTING WITHOUT EVIDENCE OF LEAKS. UNRELATED TO THE COMPLAINT, INVESTIGATION REVEALED THE DISPLAY SCREEN WAS DIM AND DISCOLORED.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (MOISTURE INGRESS) ISSUE. IT WAS REPORTED THE PATIENT¿S BLOOD GLUCOSE WAS BETWEEN 250-499 MG/DL WITHOUT SIGNS OR SYMPTOMS OF HYPERGLYCEMIA. THE REPORTED BLOOD GLUCOSE EXCURSION WITHOUT SYMPTOMS DOES NOT MEET CRITERIA FOR A SERIOUS INJURY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
412620 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 44 YR