FDA Adverse Event Malfunction Summary report: N

CAPSUREEPI

MDR report key: 3936247 · Received July 15, 2014

Report

Report Number
2182208-2014-01848
Event Type
Malfunction
Date Received
July 15, 2014
Report Date
July 14, 2014
Manufacturer
MEDTRONIC, INC.
Product Code
DTB
PMA / PMN Number
P950024
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE PROXIMAL SEGMENT WAS RETURNED, ANALYZED AND THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO A DEPRESSION WHILE IN VIVO. THE OUTER INSULATION OF THE LEAD DEVELOPED A COSMETIC DEPRESSION WHILE IN VIVO. THE OUTER INSULATION OF THE LEAD WAS OBSERVED TO HAVE BLOOD INGRESSION. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS EXPLANTED FOLLOWING A CARDIAC TRANSPLANT. THERE WERE NO PRODUCT PERFORMANCE ISSUES REPORTED WHILE IMPLANTED. THE LEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
411908 CAPSUREEPI ELECTRODE, PACEMAKER, PERMANENT DTB MEDTRONIC, INC. 4968-25

Patients

Seq Age Sex Outcome Treatment
1 00006 YR ADDR01P IPG, 4968-25 LEAD