ADVIA 1800
Report
- Report Number
- 2432235-2014-00412
- Event Type
- Malfunction
- Date Received
- July 15, 2014
- Date of Event
- June 20, 2014
- Report Date
- June 20, 2014
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- JGS
- PMA / PMN Number
- K990346
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A SIEMENS CUSTOMER SERVICE ENGINEER (CSE) WAS DISPATCHED TO THE CUSTOMER SITE. AFTER EVALUATION OF THE INSTRUMENT AND INSTRUMENT DATA, THE CSE REMOVED THE ELECTRODES TO CLEAN THE HOLDER AND TIGHTENED ALL CONNECTIONS. QUALITY CONTROLS WERE RUN, RESULTING WITHIN RANGE. THE CAUSE OF THE DISCORDANT, FALSELY ELEVATED NA RESULT IS UNKNOWN AS THE SAMPLE RESULTED AS EXPECTED UPON REPEAT TESTING ON THE SAME INSTRUMENT. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
A DISCORDANT, FALSELY ELEVATED SODIUM (NA) RESULT WAS OBTAINED ON ONE PATIENT SAMPLE ON AN ADVIA 1800 INSTRUMENT. THE DISCORDANT RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED THREE TIMES ON THE SAME INSTRUMENT, RESULTING LOWER. IT IS UNKNOWN IF THE CORRECTED RESULT WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED NA RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 412819 | ADVIA 1800 | CLINICAL CHEMISTRY ANALYZER | JGS | SIEMENS HEALTHCARE DIAGNOSTICS INC. | ADVIA 1800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |