FDA Adverse Event Malfunction Summary report: N

ADVIA 1800

MDR report key: 3936240 · Received July 15, 2014

Report

Report Number
2432235-2014-00412
Event Type
Malfunction
Date Received
July 15, 2014
Date of Event
June 20, 2014
Report Date
June 20, 2014
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC.
Product Code
JGS
PMA / PMN Number
K990346
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A SIEMENS CUSTOMER SERVICE ENGINEER (CSE) WAS DISPATCHED TO THE CUSTOMER SITE. AFTER EVALUATION OF THE INSTRUMENT AND INSTRUMENT DATA, THE CSE REMOVED THE ELECTRODES TO CLEAN THE HOLDER AND TIGHTENED ALL CONNECTIONS. QUALITY CONTROLS WERE RUN, RESULTING WITHIN RANGE. THE CAUSE OF THE DISCORDANT, FALSELY ELEVATED NA RESULT IS UNKNOWN AS THE SAMPLE RESULTED AS EXPECTED UPON REPEAT TESTING ON THE SAME INSTRUMENT. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

A DISCORDANT, FALSELY ELEVATED SODIUM (NA) RESULT WAS OBTAINED ON ONE PATIENT SAMPLE ON AN ADVIA 1800 INSTRUMENT. THE DISCORDANT RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED THREE TIMES ON THE SAME INSTRUMENT, RESULTING LOWER. IT IS UNKNOWN IF THE CORRECTED RESULT WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED NA RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
412819 ADVIA 1800 CLINICAL CHEMISTRY ANALYZER JGS SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA 1800

Patients

Seq Age Sex Outcome Treatment
1