FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3936229 · Received July 15, 2014

Report

Report Number
1416980-2014-22690
Event Type
Malfunction
Date Received
July 15, 2014
Date of Event
June 20, 2014
Report Date
June 20, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THE EVENT HISTORY LOG REVIEW SHOWED NO KEYSTROKES, PROGRAMMING, OR USE RELATED EVENTS THAT INDICATED AND/OR CONTRIBUTED TO THE REPORTED ISSUE. INTERNAL AND EXTERNAL INSPECTIONS WERE PERFORMED AND FOUND AN UNRELATED ISSUE. TEMPERATURE CONFIRMATION TESTING WAS PERFORMED AND THE PASSED. RITE (RETURNED INSTRUMENT TEST EVALUATION) ELECTRICAL TESTING WAS CONDUCTED AND THE DEVICE PASSED. THE DEVICE FAILED RITE FUNCTIONAL TESTING AS FLUID WAS TRANSFERRED ABOVE THE ALLOWABLE SPECIFICATION RANGE. A MANUAL VOLUMETRIC ACCURACY TEST WAS PERFORMED AND ALSO FAILED. FURTHER INSPECTION OF THE DOOR ASSEMBLY REVEALED AN INSUFFICIENT AMOUNT OF COPPER MESH. THE CAUSE OF THE ISSUE WAS DETERMINED TO BE THE INSUFFICIENT AMOUNT OF COPPER MESH. THE DOOR PISTON WAS SCRAPPED, AND THE DEVICE WAS SENT FOR SERVICING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE HOMECHOICE (HC) WAS RETURNED AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN DETERMINED THE HC MACHINE FAILED FLUID VOLUME ACCURACY TESTING. THE DEVICE FAILED DURING EVALUATION WITH NO PATIENT INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
412247 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1