FDA Adverse Event Malfunction Summary report: N

PROGRASP FORCEPS INSTRUMENT

MDR report key: 3936060 · Received July 15, 2014

Report

Report Number
2955842-2014-04298
Event Type
Malfunction
Date Received
July 15, 2014
Date of Event
June 17, 2014
Report Date
June 20, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS FOUND THAT THE PITCH DOWN CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE BROKEN CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE OTHER CABLES AT THE WRIST WERE UNDAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE DAMAGED CABLE, COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROGRASP FORCEPS INSTRUMENT WIRE WAS NOTED TO BE SEPARATED PRIOR TO STARTING A DA VINCI MYOMECTOMY PROCEDURE. NO FRAGMENTS FELL INTO THE PATIENT, NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
414014 PROGRASP FORCEPS INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420093-10 M10131021 648

Patients

Seq Age Sex Outcome Treatment
1