FDA Adverse Event
Malfunction
Summary report: N
ANIMAS VIBE
MDR report key: 3935974
·
Received July 15, 2014
Report
- Report Number
- 2531779-2014-20154
- Event Type
- Malfunction
- Date Received
- July 15, 2014
- Report Date
- July 7, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
Additional Manufacturer Narrative · 1
THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2014 WITH THE FOLLOWING FINDINGS: INVESTIGATION REVEALED THAT THE TEXT ON THE DISPLAY SCREEN WAS DIM, FADED, AND DISCOLORED; THE CONTRAST WAS SET TO THE MAXIMUM SETTING OF 10.
Description of Event or Problem · 1
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 414111 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |