ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-20147
- Event Type
- Malfunction
- Date Received
- July 15, 2014
- Report Date
- July 14, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1; DATE OF SUBMISSION: 09/09/2014 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION REVEALED THAT THE KEYPAD COVER WAS FREE OF DAMAGE. EVALUATION REVEALED THAT THE DOWN ARROW AND OK KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE; THE REPORTED EVENT WAS DUPLICATED. THE KEYPAD COVER WAS REMOVED, AND EVIDENCE OF CONTAMINATION WAS FOUND UNDER ALL OF THE KEY CONTACTS. UNRELATED TO THE REPORTED EVENT, THE DISPLAY SCREEN WAS DIM AND DISCOLORED. ALSO UNRELATED TO THE REPORTED EVENT, THE BATTERY COMPARTMENT WAS CRACKED.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. REPORTEDLY, THE ¿UP¿, ¿DOWN¿ AND ¿OK¿ BUTTONS WERE OVER RESPONSIVE. IT WAS REPORTED THAT THE KEYPAD COVER WAS DAMAGED AND THERE WAS NO EVIDENCE OF MOISTURE OR CORROSION IN THE PUMP. THE REPORTER DID NOT NOTE ANY DELIVERY ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 413752 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |