ENSEAL G2 CURVED JAW
Report
- Report Number
- 3005075853-2014-04967
- Event Type
- Malfunction
- Date Received
- July 15, 2014
- Date of Event
- July 9, 2014
- Report Date
- July 9, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GEI
- PMA / PMN Number
- K112033
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A. DEVICE EVALUATION - NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
(B)(4). ADDITIONAL INFORMATION: STICTION THE DEVICE WAS RECEIVED IN GOOD PHYSICAL CONDITION. IT WAS FOUND THAT THE DEVICE WAS SLOW TO RETURN TO A FULLY OPEN POSITION. THIS COULD HAVE IMPACTED THE OPENING AND CLOSING PERFORMANCE OF THE DEVICE. THE DEVICE WAS TESTED ON THE GENERATOR AND PASSED ALL FUNCTIONAL TESTING. THE ENERGY OUTPUT DELIVERED FROM THE DEVICE WAS VERIFIED AND THE CUTTING OF THE TEST MEDIA PERFORMED AS EXPECTED. ALL THREE TONES WERE HEARD DURING FUNCTIONAL TESTING (TONE 1 IS HEARD WHEN THE ENERGY ACTIVATION BUTTON IS PRESSED; TONE 2 IS HEARD WHEN TISSUE IMPEDANCE THRESHOLD IS REACHED; AND TONE 3 IS HEARD WHEN THE CYCLE IS COMPLETE). OPENING AND CLOSING OF THE JAWS MULTIPLE TIMES PRIOR TO USE WILL REDUCE OR ELIMINATE THIS ISSUE
IT WAS REPORTED THAT THE JAWS WOULD NOT OPEN DURING A TLH. PROCEDURE COMPLETED USING AN ARTICULATING NSEAL. NO PATIENT CONSEQUENCES.