SPECTRUM INFUSION PUMP
MDR report key: 3935669
·
Received February 3, 2014
Report
- Report Number
- 1314492-2014-05638
- Event Type
- Malfunction
- Date Received
- February 3, 2014
- Date of Event
- January 1, 2014
- Report Date
- January 7, 2014
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. THE DEVICE WAS FOUND OUT OF SPEC IN RELATION TO THE REPORTED SYMPTOM WHICH WAS REPRODUCED. THE DEVICE WAS TESTED FOR 24 HOURS AND NO MALFUNCTION COULD BE IDENTIFIED. SYSTEM ERROR 322 WAS CONFIRMED THROUGH AND EVENT HISTORY LOG. THE EVAL WAS UNABLE TO DETERMINE THE CAUSE. THE UPPER AND LOWER AUXILIARY ARE KNOWN CONTRIBUTORS TO THIS SYMPTOM AND HAVE BEEN REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATED AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.
Description of Event or Problem · 1
IT WAS REPORTED THAT A SPECTRUM PUMP DISPLAYED SYSTEM ERROR 322. IT WAS ALSO REPORTED THAT THERE WAS NO PT INJURY.