FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3935666 · Received February 3, 2014

Report

Report Number
1314492-2014-05632
Event Type
Malfunction
Date Received
February 3, 2014
Date of Event
January 1, 2014
Report Date
January 7, 2014
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

MFR REF NO: (B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVAL WAS PERFORMED. THE EVAL CONFIRMED AND REPRODUCED THE REPORTED SYMPTOM SYSTEM ERROR 105 CAUSED BY A FAILED MOTOR FLEX. THE MOTOR ASSEMBLY WAS REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PUMP ALARMED SYSTEM ERROR 105. IT WAS ALSO REPORTED THAT THERE WAS NO PT INJURY.