SPECTRUM INFUSION PUMP
MDR report key: 3935546
·
Received February 3, 2014
Report
- Report Number
- 1314492-2014-05893
- Event Type
- Malfunction
- Date Received
- February 3, 2014
- Date of Event
- January 1, 2014
- Report Date
- January 6, 2014
- Manufacturer
- BAXTER HEALTHCARE CORP
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
MFR REF NO.: (B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. THE DEVICE WAS FOUND OUT OF SPECIFICATION IN RELATION TO THE REPORTED SYMPTOM WHICH WAS NOT REPRODUCED. THE DEVICE WAS TESTED FOR 24 HOURS AND NO MALFUNCTION COULD BE IDENTIFIED. SYS ERROR 322 WAS CONFIRMED THROUGH THE EVENT HISTORY LOG. THE EVALUATION WAS UNABLE TO DETERMINE THE CAUSE. THE UPPER AND LOWER AUXILIARY ARE KNOWN CONTRIBUTORS TO THIS SYMPTOM AND HAVE BEEN REPLACED. SYS ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.
Description of Event or Problem · 1
IT WAS REPORTED THAT A SPECTRUM PUMP ALARMED FOR A SYS ERROR 322. IT WAS ALSO REPORTED THERE WAS NO PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 69444 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORP | 35700 |