ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-20117
- Event Type
- Malfunction
- Date Received
- July 15, 2014
- Report Date
- July 7, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/07/2014 WITH THE FOLLOWING FINDINGS: THE KEYPAD COVER WAS FOUND TO BE TORN BETWEEN THE UP ARROW BUTTON AND THE DOWN ARROW BUTTON. DURING TESTING, THE UP ARROW AND CONTRAST KEYPAD BUTTONS WERE FOUND TO BE UNRESPONSIVE. THE KEYPAD COVER WAS REMOVED AND THE UP ARROW AND CONTRAST BUTTON CONTACTS WERE FOUND TO BE MISSING. THERE WAS EVIDENCE OF CONTAMINATION FOUND UNDER THE DOWN ARROW AND OK BUTTON CONTACTS. ADDITIONALLY, THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED BETWEEN THE CASE SEAL AND GRIP PAD. FURTHERMORE, EVALUATION REVEALED THAT THE DISPLAY WAS DIM AND DISCOLORED. (B)(6).
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE UP ARROW KEYPAD BUTTON WAS UNRESPONSIVE. INVESTIGATION ALSO REVEALED THAT THE UP ARROW BUTTON CONTACT WAS MISSING AND THERE WAS CONTAMINATION UNDER THE DOWN AND OK BUTTON CONTACTS. ADDITIONALLY, INVESTIGATION REVEALED THAT THE BATTERY COMPARTMENT WAS CRACKED AND THE DISPLAY WAS DIM AND DISCOLORED. THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 07/07/2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 412638 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 23 YR |