FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3935416 · Received February 3, 2014

Report

Report Number
1314492-2014-05651
Event Type
Malfunction
Date Received
February 3, 2014
Date of Event
January 1, 2014
Report Date
January 7, 2014
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND EVALUATED. THE EVALUATION REPRODUCED THE REPORTED SYSTEM ERROR 322. PHYSICAL INSPECTION FOUND THAT THE MECHANICAL MOUNT SCREW THAT HOLDS THE UPPER AUXILIARY BRACKET WAS LOOSE ALLOWING THE UPPER LATCH SWITCH TO MOVE IN AND OUT FROM THE UPPER DOOR LATCH CAUSING IT TO BE OUT OF SPECIFICATION. THE LOOSE MECHANICAL MOUNT SCREW WILL TRIGGER AN INTERMITTENT OPEN/CLOSED SWITCH CONNECTION CAUSING THE REPORTED SYSTEM ERROR 322. THE UPPER AUXILIARY ASSEMBLY WAS REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PUMP HAS A SYSTEM ERROR 322. IT WAS ALSO REPORTED THERE WAS NO PATIENT INJURY.