FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3935415
·
Received February 3, 2014
Report
- Report Number
- 1314492-2014-05650
- Event Type
- Malfunction
- Date Received
- February 3, 2014
- Date of Event
- January 1, 2014
- Report Date
- January 7, 2014
- Manufacturer
- BAXTER HEALTHCARE CORP
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
MFR REF NO.: (B)(4). THE DEVICE WAS RETURNED TO BAXTER AND EVALUATED. THE UNIT WAS RECEIVED INOPERABLE DUE TO THE REPORTED SYSTEM ERROR 105 CAUSED BY A FAILED MOTOR (FLEX). THE MOTOR ASSEMBLY WAS REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT A PUMP HAS A SYSTEM ERROR 105. IT WAS ALSO REPORTED THERE WAS NO PATIENT INJURY.