FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3935411
·
Received February 3, 2014
Report
- Report Number
- 1314492-2014-05643
- Event Type
- Malfunction
- Date Received
- February 3, 2014
- Date of Event
- January 1, 2014
- Report Date
- January 7, 2014
- Manufacturer
- BAXTER HEALTHCARE CORP
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
MANUFACTURER REF NO.: (B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS PERFORMED. THE EVALUATION CONFIRMED THE REPORTED SYMPTOM OF SYSTEM ERROR 322 THROUGH THE HISTORY LOG, BUT COULD NOT REPRODUCE IT. REVIEW OF KNOWN CONTRIBUTORS LED TO THE DETERMINATION THAT THE UPPER AND LOWER AUXILIARY ASSEMBLIES WERE THE FAILING COMPONENTS. THE UPPER AND LOWER AUXILIARY ASSEMBLIES WERE REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET- LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.
Description of Event or Problem · 1
IT WAS REPORTED THAT A PUMP ALARMED FOR SYSTEM ERROR 322. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 71457 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORP | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |