FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3935371 · Received February 3, 2014

Report

Report Number
1314492-2014-05678
Event Type
Malfunction
Date Received
February 3, 2014
Date of Event
January 1, 2014
Report Date
January 7, 2014
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURER REF NO.: (B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS PERFORMED. THE EVALUATION CONFIRMED (THROUGH THE HISTORY LOG), BUT DID NOT REPRODUCE THE SYSTEM ERROR 322. THE DEVICE WAS OPERATING OUT OF SPECIFICATION. PAST REPAIRS HAVE SHOWN THE UPPER AND LOWER AUXILIARY ASSEMBLIES WERE REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A DEVICE IS ALARMING SYSTEM ERROR 322. IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
71691 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORPORATION 35700

Patients

Seq Age Sex Outcome Treatment
1