FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3935368 · Received February 3, 2014

Report

Report Number
1314492-2014-05670
Event Type
Malfunction
Date Received
February 3, 2014
Manufacturer
BAXTER HLTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. THE REPORTED SYSTEM ERROR 322 WAS REPRODUCED AND CONFIRMED THROUGH EVAL. PHYSICAL INSPECTION FOUND THAT THE UPPER LATCH SWITCH BRACKET SCREWS WERE LOOSE ALLOWING THE UPPER LATCH SWITCH TO MOVE IN AND OUT FROM THE UPPER DOOR LATCH. THE LOOSE SCREWS WILL TRIGGER AN INTERMITTENT OPEN/CLOSED SWITCH CONNECTION CAUSING THE REPORTED SYMPTOM. THE UPPER AUXILIARY WAS REPLACED. SYS ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SPECTRUM PUMP ALARMED SYSTEM ERROR 322. IT WAS ALSO REPORTED THERE WAS NO PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
71667 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HLTHCARE CORP 35700

Patients

Seq Age Sex Outcome Treatment
1