FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3935317 · Received July 15, 2014

Report

Report Number
1416980-2014-22632
Event Type
Malfunction
Date Received
July 15, 2014
Date of Event
January 28, 2014
Report Date
June 20, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS EVALUATED AND THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) WAS IDENTIFIED IN THE REVIEW OF THE EVENT HISTORY LOGS. THE CAUSE WAS DETERMINED TO BE A USE ERROR, INAPPROPRIATE BYPASS OF THE LOW DRAIN VOLUME ALARM, NOT INCLUDING INITIAL DRAIN. THE HOMECHOICE / HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE GIVES INSTRUCTIONS ON HOW TO BYPASS LOW DRAIN VOLUME ALARM. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, THEN A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 00:49:36. DURING NIGHT DRAIN CYCLE FIVE, THE PATIENT'S ULTRAFILTRATION READING WAS 1995ML, INDICATING THE HOME PATIENT (HP) DRAINED 1995ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO FURTHER INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
413695 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1