FDA Adverse Event Malfunction Summary report: N

LARGE NEEDLE DRIVER INSTRUMENT

MDR report key: 3935235 · Received July 15, 2014

Report

Report Number
2955842-2014-04272
Event Type
Malfunction
Date Received
July 15, 2014
Date of Event
June 4, 2014
Report Date
June 17, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION OBSERVED THAT ONE GRIP TIP WAS BENT INWARDS AT THE DISTAL END. AS A RESULT TIPS CAN NOT CLOSE COMPLETELY. EVIDENCE NOT CONCLUSIVE, BUT BENT DAMAGE MAY BE DUE TO LIKELY MISHANDLING/MISUSE. AN ADDITIONAL OBSERVATION NOT REPORTED BY SITE WAS THAT THE INSTRUMENT'S GRIP TIPS EXHIBITED CARBIDE INSERT DAMAGE. THE CARBIDE INSERT WAS BROKEN ON THE BENT TIP AND MISSING A PIECE MEASURED AT .078 LONG. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE DAMAGE TO THE INSTRUMENT'S CARBIDE INSERT FOUND DURING FAILURE ANALYSIS INVESTIGATION COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING REPROCESSING OF THE LARGE NEEDLE DRIVER INSTRUMENT, IT WAS NOTED THAT THE TIPS OF THE INSTRUMENT DID NOT MEET. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT. THERE WAS NO ALLEGATION OF ANY HARM, INJURY, OR ADVERSE OUTCOME TO A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
413493 LARGE NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420006-08 M11131203 959

Patients

Seq Age Sex Outcome Treatment
1