LARGE NEEDLE DRIVER INSTRUMENT
Report
- Report Number
- 2955842-2014-04272
- Event Type
- Malfunction
- Date Received
- July 15, 2014
- Date of Event
- June 4, 2014
- Report Date
- June 17, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION OBSERVED THAT ONE GRIP TIP WAS BENT INWARDS AT THE DISTAL END. AS A RESULT TIPS CAN NOT CLOSE COMPLETELY. EVIDENCE NOT CONCLUSIVE, BUT BENT DAMAGE MAY BE DUE TO LIKELY MISHANDLING/MISUSE. AN ADDITIONAL OBSERVATION NOT REPORTED BY SITE WAS THAT THE INSTRUMENT'S GRIP TIPS EXHIBITED CARBIDE INSERT DAMAGE. THE CARBIDE INSERT WAS BROKEN ON THE BENT TIP AND MISSING A PIECE MEASURED AT .078 LONG. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE DAMAGE TO THE INSTRUMENT'S CARBIDE INSERT FOUND DURING FAILURE ANALYSIS INVESTIGATION COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
IT WAS REPORTED THAT DURING REPROCESSING OF THE LARGE NEEDLE DRIVER INSTRUMENT, IT WAS NOTED THAT THE TIPS OF THE INSTRUMENT DID NOT MEET. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT. THERE WAS NO ALLEGATION OF ANY HARM, INJURY, OR ADVERSE OUTCOME TO A PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 413493 | LARGE NEEDLE DRIVER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420006-08 | M11131203 959 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |