FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3935161 · Received February 3, 2014

Report

Report Number
1314492-2014-05703
Event Type
Malfunction
Date Received
February 3, 2014
Date of Event
December 1, 2013
Report Date
January 7, 2014
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY BAXTER. THIS COMPLAINT WAS OPENED DURING THE INVESTIGATION OF COMPLAINT (B)(4). THE IDENTIFIED "DOWNSTREAM OCCLUSION" ALARMS WERE CONFIRMED THROUGH THE EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY ADDITIONAL INFORMATION IS RECEIVED A FOLLOW UP WILL BE SENT. THE DOWNSTREAM TUBING GUIDE, FORCE SENSOR GASKET, AND FORCE SENSOR PUSHER WERE REPLACED.

Description of Event or Problem · 1

DURING THE EVALUATION OF A RETURNED SPECTRUM INFUSION PUMP, 6 OCCURRENCES OF "DOWNSTREAM OCCLUSION" ALARMS WERE IDENTIFIED IN THE EVENT HISTORY LOG. ANY PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION IS UNKNOWN SINCE THESE ITEMS WERE FOUND DURING EVALUATION OF THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
69431 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORPORATION 35700

Patients

Seq Age Sex Outcome Treatment
1