FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3935123
·
Received February 3, 2014
Report
- Report Number
- 1314492-2014-05614
- Event Type
- Malfunction
- Date Received
- February 3, 2014
- Date of Event
- January 1, 2014
- Report Date
- January 7, 2014
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. THE DEVICE WAS FOUND OUT OF SPECIFICATION IN RELATION TO THE REPORTED SYMPTOM WHICH WAS REPRODUCED. SYSTEM ERROR 322 WAS CONFIRMED AND CAUSED BY LOOSE UPPER LATCH SWITCH BRACKET NYLON SCREWS. THIS WOULD ALLOW THE SWITCH TO MOVE CAUSING THE REPORTED SYMPTOM. THE FAILED UPPER AUXILIARY ASSEMBLY WAS REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.
Description of Event or Problem · 1
IT WAS REPORTED THAT A SPECTRUM PUMP ALARMED SYSTEM ERROR 322. IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 69472 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORPORATION | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |