FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3935102 · Received February 3, 2014

Report

Report Number
1314492-2014-05609
Event Type
Malfunction
Date Received
February 3, 2014
Date of Event
January 1, 2014
Report Date
January 7, 2014
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURER REF NO.: (B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS PERFORMED. THE EVALUATION CONFIRMED AND REPRODUCED THE "DOOR NOT FULLY LATCHED" MESSAGES CAUSED BY THE DOOR HOOKS BEING OUT OF ADJUSTMENT. THE DOOR HOOKS AND LATCH PINS WERE REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SPECTRUM PUMP WAS CONSTANTLY ALARMING "DOOR NOT FULLY LATCHED". IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
69972 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORPORATION 35700

Patients

Seq Age Sex Outcome Treatment
1