FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3935102
·
Received February 3, 2014
Report
- Report Number
- 1314492-2014-05609
- Event Type
- Malfunction
- Date Received
- February 3, 2014
- Date of Event
- January 1, 2014
- Report Date
- January 7, 2014
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
MANUFACTURER REF NO.: (B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS PERFORMED. THE EVALUATION CONFIRMED AND REPRODUCED THE "DOOR NOT FULLY LATCHED" MESSAGES CAUSED BY THE DOOR HOOKS BEING OUT OF ADJUSTMENT. THE DOOR HOOKS AND LATCH PINS WERE REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT A SPECTRUM PUMP WAS CONSTANTLY ALARMING "DOOR NOT FULLY LATCHED". IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 69972 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORPORATION | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |