FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 3935094 · Received July 15, 2014

Report

Report Number
2531779-2014-20092
Event Type
Malfunction
Date Received
July 15, 2014
Report Date
June 30, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 11/06/2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE RETURNED BATTERY CAP WAS RETURNED UNDAMAGED AND ABLE TO FIT APPROPRIATELY TO MAINTAIN AN ELECTRICAL CONNECTION. A CALL SERVICE ALARM WAS REPRODUCED UPON POWER UP AND COULD NOT BE CLEARED WITH A REBOOT. A CALL SERVICE ALARM FAILURE EVENT WAS OBSERVED IN THE BLACK BOX. COMPLETE TESTING WAS PROHIBITED DUE TO THE CALL SERVICE ALARM WHICH INDICATED THAT THERE WAS CORRUPTION IN A LANGUAGE FILE. UNRELATED TO THE INITIAL COMPLAINT, THE DISPLAY SCREEN WAS OBSERVED TO BE DIM AND REDDISH IN COLOR. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CALL SERVICE ALARM ISSUE) ISSUE. NO ADDITIONAL INFORMATION WAS AVAILABLE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
411935 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1