FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3935010 · Received July 15, 2014

Report

Report Number
2531779-2014-20036
Event Type
Malfunction
Date Received
July 15, 2014
Report Date
June 19, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 05/11/2015 WITH THE FOLLOWING FINDINGS: REVIEW OF THE BLACK BOX DATA REVEALED RECORDS OF CALL SERVICE ALARMS (054/052). ON INVESTIGATION, THE PUMP POWERED ON NORMALLY WITH A FULLY ILLUMINATED DISPLAY SCREEN. INVESTIGATION DID NOT DUPLICATE THE BLANK SCREEN COMPLAINT. UNRELATED TO THE ORIGINAL COMPLAINT, THE DISPLAY WAS NOTED TO BE DIM AND DISCOLORED.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (BLANK SCREEN) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.