FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3934930 · Received February 4, 2014

Report

Report Number
1314492-2014-05902
Event Type
Malfunction
Date Received
February 4, 2014
Date of Event
December 1, 2013
Report Date
January 8, 2014
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

MFR REF NO: (B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY BAXTER. THIS COMPLAINT WAS OPENED DURING THE INVESTIGATION OF COMPLAINT (B)(4). THE REPORTED "DOOR JAMMED" ALARMS WERE CONFIRMED THROUGH THE EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT. THE UPPER AUXILIARY SUB-ASSEMBLY, AND LOWER AUXILIARY ASSEMBLY WERE REPLACED.

Description of Event or Problem · 1

DURING THE EVAL OF A RETURNED SPECTRUM INFUSION PUMP, 2 OCCURRENCES OF "DOOR JAMMED" ALARMS WERE IDENTIFIED IN THE EVENT HISTORY LOG. ANY PT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION IS UNK SINCE THESE ITEMS WERE FOUND DURING EVAL OF THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
72559 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORP 35700

Patients

Seq Age Sex Outcome Treatment
1