MDR report key: 3934886 · Received July 15, 2014

Report

Report Number
2134265-2014-04029
Event Type
Malfunction
Date Received
July 15, 2014
Date of Event
June 16, 2014
Report Date
June 17, 2014
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
DQY
PMA / PMN Number
K053118
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4). DEVICE EVALUATED BY MFR: THE DEVICE WAS NOT RECEIVED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT BALLOON RUPTURE OCCURED. THE 100% STENOSED TARGET LESION WAS AN IN-STENT RESTENOSIS OF AN UNKNOWN STENT LOCATED IN THE MILDLY CALCIFIED AND MODERATELY TORTUOUS SUPERFICIAL FEMORAL ARTERY (SFA). A 7.0MM X 60MM X 135CM STERLING¿ BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, ON THE FIRST INFLATION AT 10 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A NON BSC BALLOON CATHETER. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
413484 STERLING? CATHETER, PERCUTANEOUS DQY BOSTON SCIENTIFIC - MAPLE GROVE H74939031706010 16176265

Patients

Seq Age Sex Outcome Treatment
1 GUIDEWIRE:TREASURE FLOPPY| INTRODUCER SHEATH:PARENT