FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3934371 · Received July 15, 2014

Report

Report Number
2032227-2014-03919
Event Type
Malfunction
Date Received
July 15, 2014
Date of Event
June 14, 2014
Report Date
June 14, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH CRACKED AND BLEEDING LCD GLASS, MINOR SCRATCHED DISPLAY WINDOW AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAS A CRACK IN THE SCREEN AND A BLACK DOT IN THE CENTER. THE BLOOD GLUCOSE READING IS 176 MG/DL. THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
412103 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAS

Patients

Seq Age Sex Outcome Treatment
1 44 YR