FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 3934348
·
Received July 15, 2014
Report
- Report Number
- 2032227-2014-03598
- Event Type
- Malfunction
- Date Received
- July 15, 2014
- Date of Event
- June 14, 2014
- Report Date
- June 14, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Additional Manufacturer Narrative · 1
NO BUTTON ERROR ALARM NOTED. INSULIN PUMP HAS INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE ON KEYPAD TRACES. INSULIN PUMP RECEIVED WITH MISSING END CAP STICKER, CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHED LCD WINDOW AND SCRATCHED RESERVOIR TUBE WINDOW.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE BLOOD GLUCOSE READING IS 83 MG/DL. THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 412027 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-523LNAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 30 YR |