FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3934323 · Received July 15, 2014

Report

Report Number
2531779-2014-20046
Event Type
Malfunction
Date Received
July 15, 2014
Report Date
June 19, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/19/2014 WITH THE FOLLOWING FINDINGS:THE PUMP POWERED ON TO A DIM, DISCOLORED DISPLAY SCREEN. ADDITIONALLY, INVESTIGATION REVEALED CONTAMINATION UNDER THE CONTRAST, UP ARROW, AND DOWN ARROW KEY CONTACTS. THE KEYPAD COVER WAS PEELING AND TORN AND THE CONTRAST KEYPAD BUTTON WAS INTERMITTENTLY UNRESPONSIVE. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE DISPLAY SCREEN WAS DIM AND THERE WAS CONTAMINATION UNDER THE BUTTON CONTACTS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 06/19/2014.