FDA Adverse Event Malfunction Summary report: N

SYSTEM 7 SAG SAW

MDR report key: 3934293 · Received July 15, 2014

Report

Report Number
0001811755-2014-02522
Event Type
Malfunction
Date Received
July 15, 2014
Date of Event
June 19, 2014
Report Date
June 19, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED DISASSEMBLY OF THE LEVER WAS CONFIRMED BY A MANUFACTURER REPAIR TECHNICIAN THROUGH VISUAL INSPECTION. POTENTIAL CAUSES FOR THE REPORTED DISASSEMBLY INCLUDE A MATERIAL FATIGUE ISSUE OR IMPACT.

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY THE DEVICE DISASSEMBLED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY, THE DEVICE DISASSEMBLED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.