SYSTEM 6 ASEPTIC HOUSING ASSY
MDR report key: 3934291
·
Received July 15, 2014
Report
- Report Number
- 0001811755-2014-02520
- Event Type
- Malfunction
- Date Received
- July 15, 2014
- Date of Event
- June 11, 2014
- Report Date
- June 19, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HWE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHARMACIST
Narratives
Additional Manufacturer Narrative · 1
THE REPORTED EVENT, DELRIN BALL IS MISSING, WAS CONFIRMED. THE TECHNICIAN OBSERVED THAT THE DELRIN BALL WAS MISSING FROM THE HOUSING. DEVICE WAS PLACED IN PARTS RETENTION.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE ASEPTIC BATTERY HOUSINGS HAVE DELRIN BALL MISSING FROM THE LATCH MECHANISM WHICH CAN LEAD TO POTENTIAL FOR HOUSING TO OPEN AND EXPOSE NON STERILE BATTERY TO THE SURGICAL SITE. IT WAS NOTICED DURING INSPECTION, NO ASSOCIATED PROCEDURE, NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.