ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-20043
- Event Type
- Malfunction
- Date Received
- July 15, 2014
- Report Date
- July 6, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/26/2014 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION OF THE PUMP FOUND SOME COSMETIC DAMAGES. REVIEW OF BLACK BOX DATA FOUND RECORDS OF THE REPORTED MOTOR RELATED CALL SERVICE ALARMS. THE PUMP POWERED UP TO THE VERIFY SCREEN WITH AUDITORY AND VIBRATORY FEATURE. ALL THE BUTTONS RESPONDED PROPERLY. DURING THE REWIND/LOAD/PRIME SEQUENCE AND A 24-HOUR PUMP EXERCISE THE PUMP EMITTED MOTOR RELATED CALL SERVICE ALARMS. THE REPORTED CALL SERVICE ALARM WAS VERIFIED IN THE BLACK BOX AND DUPLICATED IN THE INVESTIGATION. THE BATTERY COMPARTMENT WAS FOUND TO HAVE CRACKED FROM THE THREADS TO THE BASE SEAL AND MOISTURE WAS EVIDENT INSIDE THE COMPARTMENT. A LEAK TEST SHOWED A LEAK WHERE THE CRACK WAS LOCATED. THE PUMP CASING WAS OPENED TO FURTHER INVESTIGATE AND MOISTURE CONTAMINATION WAS FOUND AROUND THE MOTOR WIRE CONNECTOR ON THE PRINTED CIRCUIT BOARD.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CS 078) ISSUE. REPORTEDLY, THE PUMP ALARMED FOR MOTOR REWIND ERROR MULTIPLE TIMES IN THE PAST 30 DAYS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 413970 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR |