ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-20040
- Event Type
- Malfunction
- Date Received
- July 15, 2014
- Report Date
- July 6, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. ON INVESTIGATION, THE REPORTED CALL SERVICE ALARM WAS VERIFIED IN THE BLACK BOX BUT NOT DUPLICATED IN THE INVESTIGATION. THE REVIEW OF BLACK BOX DATA FOUND SEVERAL REPORTED CALL SERVICE 064, OCCLUSION RELATED ALARMS. THE PUMP POWERED UP WITH AUDITORY AND VIBRATORY FEATURES. NO TACTILE ISSUE WAS OBSERVED WITH THE BUTTONS. A REWIND/LOAD/PRIME SEQUENCE AND A 24-HOUR PUMP EXERCISE WERE EXECUTED WITHOUT INCIDENCES. UNRELATED TO THE REPORTED CALL SERVICE ALARM ISSUE, THE DISPLAY WAS FOUND TO BE DIM AND DISCOLORED. THE BATTERY COMPARTMENT WAS FOUND TO HAVE CRACKED FROM THE GRIP PAD TO THE CASE SEAL. THE PUMP CASING WAS OPENED AND MOISTURE CONTAMINATION WAS EVIDENT ON THE PRINTED CIRCUIT BOARD.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CS 064) ISSUE. REPORTEDLY, THE PUMP ALARMED FOR CALL SERVICE 064 AT RANDOM AND ALARM WAS NOT CLEAR BY PUMP REBOOT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 414116 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR |