FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 3934164 · Received July 15, 2014

Report

Report Number
0001811755-2014-02515
Event Type
Malfunction
Date Received
July 15, 2014
Date of Event
June 18, 2014
Report Date
June 18, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN CONFIRMED THE REPORTED EVENT OF HEATING UP AND NOTED THAT THE DRIVESHAFT BEARING WAS DAMAGED.

Additional Manufacturer Narrative · 1

THE DEVICE IS AVAILABLE FOR EVALUATION BUT HAS NOT YET BEEN RECEIVED. ADDITIONAL INFORMATION MAY BE SUBMITTED ONCE THE DEVICE IS RECEIVED AND THE QUALITY INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY THE CORE IMPACTION DRILL WAS OVERHEATING. THE CUSTOMER LET THE DEVICE COOL FOR THREE TO FOUR MINUTES BEFORE COMPLETING THE PROCEDURE SUCCESSFULLY USING THE SAME DEVICE. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY, THE CORE IMPACTION DRILL WAS OVERHEATING. THE CUSTOMER LET THE DEVICE COOL FOR THREE TO FOUR MINUTES BEFORE COMPLETING THE PROCEDURE SUCCESSFULLY USING THE SAME DEVICE. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
413701 CORE IMPACTION DRILL DRIVER, WIRE, AND BONE DRILL, MANUAL DZJ STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1