FDA Adverse Event Malfunction Summary report: N

TPS HANDPIECE CORD

MDR report key: 3934143 · Received July 15, 2014

Report

Report Number
0001811755-2014-02512
Event Type
Malfunction
Date Received
July 15, 2014
Date of Event
June 24, 2014
Report Date
June 24, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K943540
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING THE DEVICE INSPECTION, THE TECHNICIAN DETERMINED THE CABLE¿S INTERNAL WIRING WAS DAMAGED. THE DEVICE WAS DISCARDED BY THE MANUFACTURER.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING ROUTINE MAINTENANCE CONDUCTED BY A MANUFACTURER FIELD SERVICE TECHNICIAN AT THE USER FACILITY THE TPS HANDPIECE CORD CAUSED A BIAS CURRENT MESSAGE TO BE DISPLAYED ON THE CONSOLE, SIGNALING A CONDITION OCCURRED IN WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. IT WAS REPORTED THAT THERE WAS NO ASSOCIATED PROCEDURE. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
413700 TPS HANDPIECE CORD DRIVER, WIRE, AND BONE DRILL, MANUAL DZJ STRYKER INSTRUMENTS-KALAMAZOO 12338

Patients

Seq Age Sex Outcome Treatment
1