FDA Adverse Event
Malfunction
Summary report: N
TPS HANDPIECE CORD
MDR report key: 3934143
·
Received July 15, 2014
Report
- Report Number
- 0001811755-2014-02512
- Event Type
- Malfunction
- Date Received
- July 15, 2014
- Date of Event
- June 24, 2014
- Report Date
- June 24, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K943540
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AK, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DURING THE DEVICE INSPECTION, THE TECHNICIAN DETERMINED THE CABLE¿S INTERNAL WIRING WAS DAMAGED. THE DEVICE WAS DISCARDED BY THE MANUFACTURER.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING ROUTINE MAINTENANCE CONDUCTED BY A MANUFACTURER FIELD SERVICE TECHNICIAN AT THE USER FACILITY THE TPS HANDPIECE CORD CAUSED A BIAS CURRENT MESSAGE TO BE DISPLAYED ON THE CONSOLE, SIGNALING A CONDITION OCCURRED IN WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. IT WAS REPORTED THAT THERE WAS NO ASSOCIATED PROCEDURE. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 413700 | TPS HANDPIECE CORD | DRIVER, WIRE, AND BONE DRILL, MANUAL | DZJ | STRYKER INSTRUMENTS-KALAMAZOO | 12338 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |