FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3934086
·
Received February 1, 2014
Report
- Report Number
- 1314492-2014-07289
- Event Type
- Malfunction
- Date Received
- February 1, 2014
- Date of Event
- January 1, 2014
- Report Date
- January 14, 2014
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
MANUFACTURER REF NO.: (B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS PERFORMED. EVALUATION CONFIRMED THE REPORTED SYMPTOM OF SYSTEM ERROR 105 THROUGH THE HISTORY LOG, BUT COULD NOT REPRODUCE IT. FURTHER EVALUATION FOUND THAT THE MOTOR WAS INTERMITTENTLY SEIZING, WHICH IS A KNOWN CONTRIBUTOR TO SYSTEM ERROR 105. THE MOTOR ASSEMBLY WAS REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT A SPECTRUM PUMP ALARMED SYSTEM ERROR 105. IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 69398 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORPORATION | 35700 |