FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3934085 · Received February 1, 2014

Report

Report Number
1314492-2014-07287
Event Type
Malfunction
Date Received
February 1, 2014
Date of Event
January 1, 2014
Report Date
January 14, 2014
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURER REF NO.: (B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. THE DEVICE WAS FOUND OUT OF SPECIFICATION WITH RESPECT TO THE REPORTED SYMPTOM, WHICH WAS REPRODUCED. THE MESSAGES WERE FOUND TO BE CAUSED BY A FAILED LINK SWITCH ON THE UPPER AUXILIARY ASSEMBLY. THE FAILED UPPER AUXILIARY ASSEMBLY WAS REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SPECTRUM PUMP CONSTANTLY DISPLAYED THE CLOSE DOOR MESSAGE. IT WAS ALSO REPORTED THERE WAS NO PATIENT INVOLVEMENT.