FDA Adverse Event Injury Summary report: N

APEX PIN ADAPTOR - SHORT HOFFMANN LRF FOR PIN Ø3-4-5-6MM

MDR report key: 3933784 · Received July 15, 2014

Report

Report Number
0008031020-2014-00317
Event Type
Injury
Date Received
July 15, 2014
Date of Event
June 18, 2014
Report Date
June 18, 2014
Manufacturer
STRYKER TRAUMA SELZACH
Product Code
KTT
PMA / PMN Number
K113327
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE REPORTED EVENT THAT HOFFMANN LRF APEX PIN ADAPTOR HAD A STRESS CRACK CORROSION ISSUE AFTER SURGERY CAN BE CONFIRMED. THE ROOT CAUSE OF THIS BREAKAGE HAS BEEN IDENTIFIED AS A DESIGN ISSUE (MATERIAL NOT RESISTANT ENOUGH TO CORROSION). THIS PROBLEM HAS BEEN IDENTIFIED AND IS ADDRESSED. A NEW MATERIAL WILL BE IMPLEMENTED AS THIS BREAKAGE IS DUE TO A DESIGN ISSUE, A CREDIT NOTE WILL BE DONE. A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT DID NOT INDICATE ANY ABNORMALITIES. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. INDICATIONS FOR ANY MATERIAL, MANUFACTURING OR DESIGN RELATED PROBLEMS WERE NOT DETERMINED IN THE INVESTIGATION. IF ANY FURTHER INFORMATION IS PROVIDED, THE INVESTIGATION REPORT WILL BE UPDATED.

Description of Event or Problem · 1

SHORT APEX PIN ADAPTOR SNAPPED. PATIENT HAD FRAME ON 3 MONTHS BEFORE IT SNAPPED.

Description of Event or Problem · 1

SHORT APEX PIN ADAPTOR SNAPPED. PATIENT HAD FRAME ON 3 MONTHS BEFORE IT SNAPPED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
412413 APEX PIN ADAPTOR - SHORT HOFFMANN LRF FOR PIN Ø3-4-5-6MM APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT KTT STRYKER TRAUMA SELZACH Z12644

Patients

Seq Age Sex Outcome Treatment
1 55 YR Required Intervention