CODMAN® MALIS® INTEGRATED TUBE SET UNITIZED PLUG
Report
- Report Number
- 1226348-2014-11812
- Event Type
- Malfunction
- Date Received
- July 15, 2014
- Date of Event
- June 18, 2014
- Manufacturer
- CODMAN & SHURTLEFF
- Product Code
- GEI
- PMA / PMN Number
- PK890648
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE MALIS TUBING SET WAS RETURNED UNOPENED AND IN THE ORIGINAL PACKAGING. PRODUCT CODE IS 80-1189 (MALIS TUBING SET). THE LOT CODE PROVIDED WAS 449672. SINCE THE LOT NUMBER WAS PROVIDED, THE PRODUCTION RECORD FOR THE RETURNED ITEM WAS REVIEWED. THE RECORD WAS COMPLETE AND NO NON-CONFORMANCES WERE FOUND. THE PRODUCT WAS RE-INSPECTED AND DID NOT PASS THE VISUAL AND TAPPI INSPECTION REQUIREMENTS. IN REVIEWING THE RETURNED ITEM, FOREIGN MATTER (LIKE A HAIR) WAS FOUND IN THE OUTER BARRIER FOR THE UNOPENED PACKAGE. THE EXACT ROOT CAUSE COULD NOT BE DETERMINED. THE MALIS TUBING SET WAS RETURNED UNOPENED AND IN THE ORIGINAL PACKAGING. SINCE THE LOT NUMBER WAS PROVIDED, THE PRODUCTION RECORD FOR THE RETURNED ITEM WAS REVIEWED. THE RECORD WAS COMPLETE AND NO NON-CONFORMANCES WERE FOUND. THE OPERATORS WERE RE-TRAINED. ANY ADDITIONAL COMPLAINTS FOR MALIS TUBING SETS WILL BE MONITORED. NO FURTHER ACTIONS ARE REQUIRED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME THIS COMPLAINT IS CONSIDERED CLOSED.
UPON COMPLETION OF INVESTIGATION, A FOLLOW UP REPORT WILL BE FILED. DEVICE IS UNDER INVESTIGATION.
FOREIGN MATTER (HAIR) WAS FOUND IN THE PACKAGE BEFORE USE. USER TRAINING SCHEDULE IS UNKNOWN. THERE WAS NO SURGICAL DELAY. THIS IS NOT THE PATIENT RELATED MATTER WHICH OCCURRED BEFORE THE SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 412082 | CODMAN® MALIS® INTEGRATED TUBE SET UNITIZED PLUG | IRRIGATION TUBING SET, GENERAL PURPOSE | GEI | CODMAN & SHURTLEFF | 449672 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |