FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3933497 · Received July 14, 2014

Report

Report Number
2032227-2014-03849
Event Type
Malfunction
Date Received
July 14, 2014
Date of Event
June 14, 2014
Report Date
June 14, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP RECEIVED WITH BUTTON ERROR ALARM AND HAD NO BUTTON RESPONSE DUE TO MOISTURE DAMAGE ON KEYPAD TRACES. UNABLE TO PERFORM DISPLACEMENT TEST DUE TO NO BUTTON RESPONSE. THE INSULIN PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON LCD WINDOW, SCRATCHED RESERVOIR TUBE WINDOW AND STAINED END CAP STICKER.

Description of Event or Problem · 1

CUSTOMER REPORTED THE INSULIN PUMP ALARMED BUTTON ERROR. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 88 MG/DL. CURRENTLY DOES NOT RECALL ANY SIGNIFICANT EVENT THAT MAY HAVE CAUSE THE DEVICE TO ALARM. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
411228 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523LNAL

Patients

Seq Age Sex Outcome Treatment
1 65 YR