FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3933379 · Received July 14, 2014

Report

Report Number
2531779-2014-19996
Event Type
Malfunction
Date Received
July 14, 2014
Report Date
July 4, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/29/2014 WITH THE FOLLOWING FINDINGS: THE PUMP WAS RETURNED WITH VISIBLE EVIDENCE OF MOISTURE BEHIND THE DISPLAY SCREEN. DURING TESTING, THE DISPLAY SCREEN WAS FOUND TO BE CLOUDY AND DIFFICULT TO READ DUE TO MOISTURE. THE PUMP WAS OPENED AND MOISTURE DAMAGE WAS FOUND ON THE INTERNAL COMPONENTS OF THE PUMP. UNRELATED TO THE INITIAL COMPLAINT, THE PUMP CASE WAS FOUND TO BE CRACKED. THE LEAK TEST FAILED DUE TO A CRACKED PUMP CASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (CLOUDY W/ MOISTURE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.