FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3933378 · Received July 14, 2014

Report

Report Number
2531779-2014-19999
Event Type
Malfunction
Date Received
July 14, 2014
Report Date
July 5, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON 08/21/2014 WITH THE FOLLOWING FINDINGS: THE PUMP HISTORIES COULD NOT BE DOWNLOADED DUE TO MOISTURE INGRESS. THE PUMP POWERED ON PROPERLY, AND THE DISPLAY SCREEN WAS FUNCTIONING; THE REPORTED EVENT WAS NOT DUPLICATED. THE PUMP FAILED A LEAK TEST DUE TO A BATTERY COMPARTMENT LEAK. THE PUMP CASE WAS REMOVED, AND EVIDENCE OF MOISTURE INGRESS WAS OBSERVED ON INTERNAL COMPONENTS.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY BLANK SCREEN W/MOISTURE IN THE BATTERY COMPARTMENT ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
409455 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 40 YR