FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3933299 · Received July 14, 2014

Report

Report Number
3004209178-2014-87044
Event Type
Malfunction
Date Received
July 14, 2014
Date of Event
June 15, 2014
Report Date
June 15, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSULIN PUMP HAD A CRACKED RESERVOIR TUBE LIP AND MINOR SCRATCHES ON DISPLAY WINDOW NOTED DURING VISUAL INSPECTION. NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED.

Description of Event or Problem · 1

CUSTOMER CALLED IN REPORTING THAT HER INSULIN PUMP IS ALARMING BUTTON ERROR. CUSTOMER'S BLOOD GLUCOSE LEVEL IS 168 MG/DL. THE DEVICE DID NOT GET WET, BUT IT MAY HAVE BEEN EXPOSED TO SWEAT. CUSTOMER WAS ADVISED THAT THE DEVICE NEEDS TO BE REPLACED, AND TO DISCONTINUE ITS USE AND REVERT TO A BACK UP PLAN. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
409801 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAH

Patients

Seq Age Sex Outcome Treatment
1 42 YR